CombiFlash Torrent® User Manual
3-16
When the Stop button is pressed, solvent remains in the tubing
from the injection valve to the detector’s flow cell. Follow the steps
listed below, according to your instrument’s configuration.
• Solid Sample Cartridge and air purge is enabled:
If you suspend the run with the Stop button, do not remove
the column and/or solid sample cartridge without first
pressing the Fast Forward button to purge solvents from all
lines. If you stop the run because your peak of interest has
been collected, protect the next user from being sprayed by
solvents left in the lines by pressing Fast Forward so the
instrument purges itself and is safe for its next use.
• Solid Sample Cartridge and air purge is NOT enabled,
i.e. purge time set to zero in Method Editor:
If you suspend the run with the Stop button, remove the
column and install the Prime Tube in place of the column.
Manually purge the Sample Cartridge by pressing Tools ->
Manual Control -> Air Purge Cartridge. Repeat this
procedure for Air Purge Column.
• Liquid sample is loaded onto column and air purge is
enabled:
If you suspend the run with the Stop button, remove the
column and install the Prime Tube in its place. Press Fast
Forward to purge all liquid lines.
• Liquid sample loaded onto column with air purge NOT
enabled:
If you suspend the run with the Stop button, remove the
column and install the Prime Tube in its place. Manually
purge the liquid lines by pressing Tools -> Manual Control
-> Air Purge Column.
• Next Tube — This button advances the fraction
collector/fractionation valve to the next tube position.
This allows you to conveniently collect eluate of interest
in a new tube.