CombiFlash Torrent® User Manual
4-28
Enable solvent B priming pump — The CombiFlash
Torrent module is equipped with a priming pump for the
B solvent to ensure that the system forms repeatable
gradients. This pump can be enabled or disabled.
CombiFlash Torrent Name Enter a name for the module. This
name will appear in the title bar of the Main window. The
name will also appear in run summaries. Unique names are
helpful when operating more than one CombiFlash Torrent
module in your lab.
Time Zone Select your local time zone.
Set Date/Time Click this button to set the date and time. From
the touch screen, this command opens a window that allows
you to set the instrument date and time. If you select this
command from a remotely connected computer, this menu
command automatically sets the module time to match the
computer’s.
Note: The module will restart after you set the time.
To prevent unauthorized time changes, this feature is
password protected. See also: S
ET PASSWORD [4.3.11].
Vapor Limit Type or select the vapor limit setting. The
CombiFlash Torrent module’s internal vapor sensor detects
vapors present in the module. When this value is exceeded,
the module will shut down to avoid a hazardous condition. For
more information, refer to S
ETTING THE VAPOR LIMIT [4.4.11].
Set Default Tube Volume Click this button to open a window
from which you can set the default volume of the collection
tubes. You can increase or decrease these values to set the
desired default volume collected in each container. Ensure
that you do not enter a value that exceeds the container’s
capacity. This will overfill the container.
The Foxy fraction collector will automatically advance to the
next tube when this default volume is met in each tube. Keep
in mind that the P
EAK COLLECTION [4.3.2.5] settings on the
M
ETHOD EDITOR [4.3.2] window may override this default
Vista de pagina 107