CombiFlash Torrent® User Manual
2-10
If using the optional Waste Outlet Cap (60-5394-512 for 83B
closures, 60-5394-513 for 100 mm closures), proceed with the
following steps:
3. Place the metal plate inside the plastic cap (Figure 2-2).
The plate should be oriented so that the small tube will
extend into the carboy. The small tube should extend 5 ±0.5
cm long from nut. If the tube is missing, insert it now and
finger-tighten the nut.
Figure 2-2 Optional Waste Cap (83B cap shown)
4. The tubing is secured using metal, spring-type retainers. A
pair of conventional pliers will allow you to open the
retainer enough to slide it onto the tubing. On the larger
diameter waste tube, slide one retainer about 2 inches (5
cm) onto the tubing end.
5. Feed this tube into the top of the metal plate until the
retainer is against the top of the plate.
6. Slide a retainer onto the tubing end until it is seated
against the bottom of the metal plate.
7. Thread the cap onto the carboy.
8. The smaller diameter tube is the level sense tube. Push
this tube about 2 cm into the top fitting and finger-tighten
the nut.
Nut
Metal Plate
5 cm tube extending
into carboy
Spring-type
Retainer
Place a retainer above and
below the metal plate.
Comentarios a estos manuales