CombiFlash Torrent® User Manual
4-22
To open the Gradient Optimizer window, open the METHOD
E
DITOR [4.3.2] window, then click the Gradient Optimizer
button.
4.3.4 Column Data
The Column Data window reports information about the
RediSep Rf column installed in the module. Compatible
RediSep Rf and RediSep Rf Gold columns have an RFID tag
that identifies the column type. PeakTrak reads this
information to identify the size and media, then automatically
loads a default method for the inserted column.
If the column is reusable, the module will also write
information to the RFID tag that can help you manage the
columns use. TheNumber of times used,First used on,” and
“Last used on” information can help you determine when the
column should be replaced. The “Last fluid used” will help you
determine if any solvent remaining in the column will be
miscible with the solvent currently used in the module.
To view the Column Data window, open the M
ETHOD EDITOR
[4.3.2] window, then click the Column Data button.
4.3.5 Files
The Files window is adaptive. That is, its function and
features change according to the command used to open the
window. Menu commands such as F
ILE>OPEN [4.2.1.2] and SAVE
M
ETHOD AS[4.2.1.6], or clicking Open and Save As buttons
will open this window. Use this window to browse the module’s
files and folders. The following controls appear on the window:
Current Path — The top-left corner of the window
displays the path (current folder). As you browse
through the files, the path will update as you go. You can
click the folder names to return to upper folder levels.
File and Folder Operation buttons:
Copy — Click this button to copy a highlighted file to
the module’s clipboard memory.
Paste — Click this button to paste a file from the
clipboard memory. If the file already exists in the
current folder, the module will ignore the Paste
Vista de pagina 101