CombiFlash Torrent® User Manual
2-20
2.10.3 Optional Fractionation Valve
The fractionation valve is an alternative to the typical Foxy
fraction collector setup, allowing the effluent from the Torrent
to be collected in any of six user-supplied containers. The
fractionation valve module includes a diverter valve to allow
the effluent to be diverted to waste.
Figure 2-11 Optional fractionation valve
1. Position the Fractionation Valve on top or near the Torrent.
If positioning the valve on the top of the Torrent, attach the
supplied mounting plate to the Fractionation Valve’s front
panel using two panhead screws. Next, position the valve
on top of the Torrent and align the mounting plate with the
holes at the right side panel (top). Secure the valve to the
module using two panhead screws.
Alternatively, the valve can be mounted on a mast.
Figure 2-12 Fractionation valve mounting holes on Torrent