Asylum AFM SOP
1
Model:
Serial Number:
MFP Controls
Before flipping over Housing, raise the stage to its MAX as to not crash the tip into the surface
1. Flip over Housing; make sure Housing is level
2. Click Live Video (Screen Image on bottom)
3. Turn on light at computer. Align at instrument to find the cantilever
4. Adjust back & right dials to increase SUM to its MAX VALUE
5. Adjust left dial to make REFLECTION = 0
6. Turn light off, close live video
Tune Tab
1. Auto Tune High = 400 KHz (Vistaprobe) / 100 KHz (Asylum)
2. Target % = -5%
3. Click Auto Tune
Main
1. Scan Size = 50 μm (can vary)
2. Scan Points = 256
3. Scan Lines = 256
4. Set Point = 900 (or 800)
5. Data Scale = 400 nm
Meter
1. Click Engage
2. AMPLITUDE in range of 1
3. Z VOLTAGE 150
4. Bring down tip
5. AMPLITUDE gets to 0.90 (Ding!)
6. Go down until Z VOLTAGE = 30-40
7. Close Housing
Main
1. Click Set Point Dot (Activates Wheel)
1
9/15/2010