Concentration range (Low to High)
Recording range (Low to High)
Sensitivity
Auto scan (if necessary)
Response time (0.02)
Repetitions
If using Mulitpoint Curve, select order of curve and zero intercept
Select OK to validate
Prepare standards (at least 4).
Auto zero the instrument with sample compartment empty or with a cuvette with solvent used as
a blank.
Select STANDARD push button.
Place first standard in cuvette and put into the sample compartment. Select READ. The EDIT
STANDARD box appears. Enter the appropriate data for the concentration and select OK.
Repeat for each standard.
Place unknown in cuvette and place in sample compartment. Select the UNKNOWN button and
then the READ button for each unknown.
When completed reading unknowns, select FILE, Save Channel. Click on the STANDARD box
and type in a file name. Also type in comments if necessary. This will be saved as a .std file (do
not type in the extension).
Click on the UNKNOWN box and enter a file name (do not enter an extension). This will be
saved as a .qnt file.
Select OK to save files.
TIME COURSE
Acquire menuTime Course
Set parameters
EX / EM wavelength
Recording range (low to high)